首页 | 本学科首页   官方微博 | 高级检索  
     


Preconcentration technique for the determination of trace elements in natural water samples by ICP-AES
Authors:S. Hari Babu  K. Suresh Kumar  K. Suvardhan  K. Kiran  D. Rekha  L. Krishnaiah  K. Janardhanam  P. Chiranjeevi
Affiliation:(1) Environmental Monitoring Laboratory, Department of Chemistry, S. V. University, Tirupati, 517502, India;(2) Department of Environmental Sciences, S. V. University, Tirupati, 517502, India
Abstract:A procedure was developed for the determination of Cd, Cu, Zn, Co, Ni, Mn, Pb and Mo in water samples by inductively coupled plasma atomic emission spectrometry (ICP-AES) after preconcentration on a morpholine dithiocarbamate (mor-DTC) supported by bagasse (Saccharam aphisinaram). The sorbed elements were subsequently eluted with 4,M HNO3 and the acid eluates were analysed by ICP–AES. The influence of various parameters such as pH, flow rate of sample, eluent concentration, volume of the sample and volume of eluent were investigated to enhance the sensitivity of the present method. A 20,mL disposable syringe served as preconcentration column. Under the optimal conditions Cd, Cu, Zn, Co, Ni, Mn, Pb and Mo in aqueous sample was concentrated about 100-fold. The sorption recoveries of elements were higher than 99.6%. The method is also applied for the analysis of natural and spiked water samples.
Keywords:Trace elements  Preconcentration  Morpholine dithocarbamate (mor-DTC)  Inductive coupled plasma –   atomic emission spectrometry  Bagasse and water samples
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号