首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Measurement of Volatile Organics at Part per Billion Concentrations Using a Cold Trap Inlet and High Speed Gas Chromatography
Authors:Robert F Mouradiarn  Steven P Levine  Hui-Qiong Ke  Herbert H Alvord
Institution:The University of Michigan School of Public Health, Department of Environmental and Industrial Health , Ann Arbor , Michigan , USA
Abstract:Direct inlet gas chromatography (GC) is becoming one of the most frequently used techniques for measurement of volatile organics in air. Although GC is an effective tool, its usefulness as a field method can be limited by retention times of several minutes or more, and by the limits of detection. In order to address these issues, a high speed GC system featuring a capillary cold trap inlet and variable speed electrometeramplifier was developed and tested.

The gas cooled inlet was used to collect and focus organic vapors from injection volumes of up to 1 ml. Resistance heating of the metal cold trap produced a narrow injection band that allowed simple separations to be completed in 5 to 20 seconds. Use of a variable speed electrometer-amplifier allowed the response time and noise suppression characteristics to be adjusted to the needs of the specific analysis.

Simple mixtures of organics including aromatics, alkanes and chlorinated hydrocarbons were separated in 20 seconds or less. The limits of detection for 13 test compounds ranged from less than 1 part per billion by volume (ppbv) to about 50 ppbv. Data presented here indicate that high speed GC using a cold trap inlet may be a feasible approach for near real-time measurement of volatile organics in ambient air.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号