首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A method for determining pyrene in mucus using synchronous fluorimetry with multiple standard additions
Authors:Xie Yanjiao  Zhu Ying  Xu Fu-Liu  Liu Xianli  Xue Bei  Tao Shu
Institution:College of Environmental Sciences & MOE Laboratory for Earth Surface Process, Peking University, Hai-Dian, Beijing 100871, PR China.
Abstract:A new method was proposed to determine pyrene in mucus, which combined the synchronous fluorimetry with the multiple standard addition method (SFMSA). The method was used to determine pyrene in mucus directly without pretreatment. The method detection limit (MDL) for pyrene in mucus was measured as 0.47 ng/ml with a relative standard deviation of 12.7% (n = 7). The standard addition graph was linear in the range 0.05-50.00 ng/ml (r(2) = 0.9989). SFMSA was validated using a GC/MS method as a reference method, and nice agreement was found. The pyrene in mucus can be directly monitored by SFMSA without solvent extraction of samples. This indicates that SFMSA is more timesaving, less laborious and cheaper than the GC/MS method with solvent extraction. SFMSA has lower MDL and higher average recovery than the GC/MS method.
Keywords:
本文献已被 PubMed 等数据库收录!