A Cyclone for Size-Selective Sampling of Ambient Air |
| |
Authors: | Walter John Georg Reischl |
| |
Institution: | California Department of Health Services , Berkeley , California , USA |
| |
Abstract: | A cyclone with a 47 mm after-filter has been developed for ambient air size-selective monitoring. It has been extensively evaluated with laboratory-generated aerosol. Variation of the pressure drop and 50% cut point with flow rate show that the cyclone operates in a single flow regime with a vortex in the outlet flow. The particle size cutoff curve is comparable in sharpness to a cascade impactor and is the same for solid or liquid particles. At 21.7 L/min, D 50 is 2.5μm and at 15.4 L/min, D 50 is 3.5 μm. Collection efficiency data for flow rates from 8 to 27 L/min fit a universal curve when plotted vs. the normalized particle diameter, (D-D 50)/D 50 Reentrainment of previously deposited particles is less than 1 % of the loading per day. In field tests the cyclone has proved to be a very satisfactory size-selective sampler. |
| |