A Portable,Battery-Operated,Immediate-Readout Dust-Particle Analyzer
Authors:V Withstandley  WJ Moroz  GW Anderson
Institution:The Pennsylvania State University
Abstract:A portable, battery-operated, electro-optical dust analyzer operating on the principle of light scattering from individual dust particles in air drawn through a high-intensity light field has been developed for monitoring dust levels where external power may be unavailable. The instrument counts and size-discriminates airborne particles over two size ranges upwards of 0.3 microns with immediate readout. A dilution mechanism permits analysis of particulate concentrations as high as 2400 particles/cm3 on each channel. The self-contained Ni-Cd batteries are rechargeable. Total weight, including batteries and all components, is approximately 14 Ib. Total power used, mainly for pumping a filtered air sheath at 300 cm/sec and for the light source, is less than 34 watts. Another unique feature is the flow system designed for turbulence suppression.

Electronic calibration in terms of number of particles/cm3 for each size range is accomplished in the laboratory by means of a pulse generator with respect to which the readout indicator is calibrated for count. The input voltages to the pulse-height discriminators are appropriately attenuated for various pulse amplitudes simulating the output of the light-sensitive pickup, a photomultiplier tube.Pulse height as a function of particle size was measured for polystyrene latex beads ranging from 0.3 to 2 microns in diameter. Good correlation is obtained with results from commercial analyzers.

Field surveys have been undertaken with the CAES prototype analyzer at various sites in central Pennsylvania and at numerous sampling points within the cities of Pittsburgh and Philadelphia. Particulate concentrations as high as 1400 particles/ cm3 have been measured at a coal cleaning plant.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号