Development of an efficient and sensitive analytical method for the determination of copper at trace levels by slotted quartz tube atomic absorption spectrometry after vortex-assisted dispersive liquid-liquid microextraction in biota and water samples using a novel ligand
Authors:Buse Tu?ba Zaman  Emine Gülhan Bak?rdere  Nursu Aylin Kasa  Serenay Deniz  Sabriye Sel  Dotse Selali Chormey  Sezgin Bak?rdere
Institution:1.Chemistry Department, Faculty of Art and Science,Y?ld?z Technical University,?stanbul,Turkey;2.Department of Science Education, Faculty of Education,Y?ld?z Technical University,?stanbul,Turkey
Abstract:This study describes the determination of trace levels of copper by slotted quartz tube atomic absorption spectrometry after dispersive liquid-liquid microextraction. A ligand synthesized from the reaction of salicylaldehyde and 1-naphthylamine was used to form coordinate copper complex prior to extraction. All parameters that influence the output of complex formation, extraction, and instrumental measurement were optimized to enhance the absorbance signal of copper. Under the optimum conditions, about 104-fold enhancement in sensitivity was recorded over the conventional flame atomic absorption spectrometer, corresponding to a 0.51 ng/mL detection limit. The percent relative standard deviation calculated for the lowest concentration (4.8%) indicated high precision for the experimental procedure. Accuracy and applicability of the optimum method were determined by performing spiked recovery tests on urine, lake water, and mineral water samples. Satisfactory recovery results were obtained between 82.2 and 106.3% at four different concentrations. Matrix matching method was also performed to increase the accuracy of quantification, and the percent recovery calculated for 175 ng/mL was 105.14%.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号