首页 | 本学科首页   官方微博 | 高级检索  
     


Measurements in the raw gas of a full scale municipal waste incinerator using a wavelength resolved REMPI mass spectrometer
Authors:Nomayo M  Thanner R  Pokorny H  Grotheer H H  Stützle R
Affiliation:DLR Institute of Combustion Technology, Stuttgart Paffenwaldring 38, Germany.
Abstract:A resonance enhanced multiphoton ionization (REMPI) mass spectrometer consisting of a tunable laser unit (Nd:YAG pump laser and OPO) and a reflectron mass spectrometer have been upgraded to cope with the difficult conditions such as vibrations, temperature fluctuations and dust, prevailing in an incinerator. On-line measurements of the raw gas have been carried out at the municipal waste incinerator in Stuttgart. Although this test series could not be completed the results are encouraging. They show that wavelength resolution can provide valuable information beyond the one obtainable by fixed frequency REMPI. Examples discussed include separation of interferants to the aniline signal and identification of phenantrene.