直读光谱仪碳元素检出限不确定度分析
作者姓名:张令莉  陈瑶
作者单位:[1]大庆市计量检定测试所 [2]齐齐哈尔市环境监测中心
摘    要:依据JJG1059-1999对直读光谱仪碳元素检出限的不确定度进行了评估,分析了影响测量不确定度的各个因素,计算得出直读光谱仪碳元素检出限的扩展不确定度为:U=0.11%,k=2。

关 键 词:直读光谱仪    检出限  不确定度

Analysis on Uncertainty of Detection Limit of Carbon with Direct Reading Spectrometer
Authors:ZhangLingliMetrological Verification Testing Institute of DaQing City; DaQing HeiLongjiang  ChenYao
Institution:(Environment Monitoring Center Station of Qiqihaer City; Qiqihaer HeiLongjiang 161005);
Abstract:Based on JJG1059-1999,the uncertainty of detection limit of carbon with direct reading spectrometer was assessed.A,the same time,the various factors which effect the uncertainty of measurement were analyzed. Then,the expanded uncertainty was obtained by calculation:u = 0.0011%,k = 2.
Keywords:Direct reading spectrometer  Carbon  Limit of detection  Uncertainty