首页 | 本学科首页   官方微博 | 高级检索  
     检索      

程序控温石墨消解-原子荧光法测定土壤中总砷
引用本文:吕睿.程序控温石墨消解-原子荧光法测定土壤中总砷[J].环境监测管理与技术,2014,26(6):50-52.
作者姓名:吕睿
作者单位:甘肃省环境监测中心站,甘肃 兰州,730020
摘    要:采用程序控温石墨全自动消解仪消解土壤样品,原子荧光光谱法测定总砷。通过试验确定原子荧光光度计的最佳工作条件,使得该方法在2.00μg/L ~20.0μg/L 范围内线性良好,检出限为0.025 mg/kg,加标回收率在96.0%~10^5%之间,RSD为1.9%~3.3%。用该方法与国标法同时测定土壤标准样品和实际样品,结果无显著差异。

关 键 词:总砷  程序控温石墨消解  原子荧光法  土壤

Determination of the Total Arsenic in Soil by Atomic Fluorescence Spectrometry with Program Temperature Controlled Graphite Digestion
LV Rui.Determination of the Total Arsenic in Soil by Atomic Fluorescence Spectrometry with Program Temperature Controlled Graphite Digestion[J].The Administration and Technique of Environmental Monitoring,2014,26(6):50-52.
Authors:LV Rui
Institution:LV Rui (Gansu Province Environmental Monitoring Center Station, Lanzhou, Gansu 730020, China)
Abstract:Program temperature controlled graphite digestion method was developed and used in the determi-nation of the total selenium in soil by atomic fluorescence spectrometry (AFS).The digestion procedure and the optimal working conditions were found to obtain the satisfactory results.The linear range was 2.00 μg/L to 20.0 μg/L,the detection limit was 0.025 mg/kg,the recoveries were from 96.0% to 10^5%,and the relative standard deviation ranged from 1.9% to 3.3%.Using the current method and the national standard method to determine standard soil sample and actual soil sample simultaneously,the results showed no significant differ-ence.
Keywords:Total Arsenics  Program temperature controlled graphite digestion  Atomic fluorescence spectrometry  Soils
本文献已被 维普 万方数据 等数据库收录!
点击此处可从《环境监测管理与技术》浏览原始摘要信息
点击此处可从《环境监测管理与技术》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号