引用本文:周 雯,史啸勇.Smart现场分析仪测定化学需氧量、氨氮[J].环境监测管理与技术,2000,12(1):33-34.
作者姓名:周 雯  史啸勇
作者单位:江苏省环境监测中心!江苏南京210029
摘    要:对Smart现场分析仪测定化学需氧量、氨氮作了简单地介绍,并用实际水样与化学法进行了对比试验。测定结果表明,两者之间相对误差均在11%以内,较为一致。仪器体积小、携带方便,操作简单、快捷,试剂消耗少,可在现场从事应急监测。

关 键 词:现场分析仪  化学需氧量  氨氮  水质监测

Determination of COD and NH 3 -N by Smart PortableInstrument
ZHOU Wen,SHI Xiao yong.Determination of COD and NH 3 -N by Smart PortableInstrument[J].The Administration and Technique of Environmental Monitoring,2000,12(1):33-34.
Authors:ZHOU Wen  SHI Xiao yong
Abstract:Determination of COD and NH 3 N by Smart portuble instrument was briefly introduced, along with compare between chemical method and this method of analysis of real samples.The results showed the relative error was no more than 11%. The instrument is small, simple and fast in operation, less consumption of reagents. It is useful for accidental monitoring.
Keywords:Portable  instrument  COD  NH  3  N  