A diffusive sampling device for the determination of formaldehyde in air using N-methyl-4-hydrazino-7-nitrobenzofurazan (MNBDH) as reagent |
| |
Authors: | Büldt A Lindahl R Levin J O Karst U |
| |
Institution: | Westf?lische Wilhelms-Universit?t Münster, Anorganisch-Chemisches Institut, Abteilung Analytische Chemie, Wilhelm-Klemm-Str. 8, 48149 Münster, Germany. |
| |
Abstract: | A new method utilizing the diffusive sampling of formaldehyde in air has been developed. Formaldehyde is sampled with the use of a glass fiber filter impregnated with N-methyl-4-hydrazino-7-nitrobenzofurazan (MNBDH) and phosphoric acid. The formaldehyde hydrazone formed is desorbed from the filter with acetonitrile and determined by high-performance liquid chromatography (HPLC) with UV/visible detection at 474 nm. The sampling rate was determined to be 24.7 mL min-1 with a relative standard deviation of 7% for 48 experiments. The measured sampling rates were not dependent on the formaldehyde concentration (0.1-1.0 mg m-3), sampling time (15-482 min) or relative humidity (20-85%). The detection limit was 70 micrograms m-3 for a 15 min sampling period and 2 micrograms m-3 for an 8 h sampling period. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|