首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Validation of a diffusive sampler for NO2
Authors:Hagenbjörk-Gustafsson A  Lindahl R  Levin J O  Karlsson D
Institution:National Institute for Working Life, Department of Chemistry, P.O. Box 7654, S-907 13 Ume?, Sweden.
Abstract:A diffusive sampler for NO2, Willems badge, was validated in laboratory experiments and field tests. The collecting reagent for NO2 in the sampler is triethanolamine, and the analysis is based on a modified colorimetric method, the Saltzman method. The analysis was performed by a flow injection analysis (FIA) technique. The sampling rate for the sampler was determined to be 40.0 ml min-1. There was no effect of NO2 concentration or relative humidity on sampling rate, and the influence of sampling time was found to be small. The detection limit was 4 micrograms m-3 for a 24 h sample. The capacity is high enough to allow sampling of 150 micrograms m-3 for 7 days, which is twice the recommended Swedish short-term (24 h) guideline value as a 98-percentile over 6 months. In field tests, the sampler performed well, even at wind speeds higher than 2 m s-1, and at low temperatures. The overall uncertainty of the method was 24%. The sensitivity and capacity of the method also make it suitable for personal sampling for 2-8 h in working environments.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号