首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Trace level determination of perchlorate in water matrices and human urine using ESI-FAIMS-MS
Authors:Ells B  Barnett D A  Purves R W  Guevremont R
Institution:Institute for National Measurement Standards, National Research Council of Canada, Ottawa, ON.
Abstract:High-field asymmetric waveform ion mobility spectrometry (FAIMS) separates perchlorate from interfering isobaric ions of bisulfate and dihydrogenphosphate in the gas-phase. The use of a new FAIMS prototype and waveform generator, along with the use of a mixed carrier gas, in this electrospray ionisation (ESI)-FAIMS-mass spectrometry (MS) study gave a detection limit for perchlorate in a relatively "clean" matrix of tap water of 0.050 ppb. Flow injection analysis (FIA) of dilutions of fortified waste water, a fortified river water certified reference material (CRM; SLRS-4, National Research Council of Canada), and a fortified human urine Standard Reference Material (SRM; 2381, National Institute of Standards and Technology) gave detection limits of 0.37 ppb, 0.50 ppb, and 4.8 ppb, respectively, in the undiluted matrices.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏