首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Comparison of collection efficiencies of sampling methods for removable beryllium surface contamination
Authors:Dufay Sarah K  Archuleta Melecita
Institution:The Plus Group, Sandia National Laboratories, PO Box 5800 MS 1103, Albuquerque, New Mexico 87185-1103, USA. skdufay@yahoo.com
Abstract:The primary goal of this study is to compare wet versus dry sampling methods for three different media, (Whatman 41 filter papers, GhostWipes, and SKC Smear Tabs), These media were selected because they are the most commonly used means for the collection of beryllium from contaminated surfaces within the DOE Complex. A range of known concentrations of beryllium was introduced onto a smooth nonporous surface. All three types of media were tested for collection efficiency in both the wet and dry mode and analyzed by inductively-coupled plasma atomic emission spectrometry (ICP-AES).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏