首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of applicability of the Sartorius Airport MD8 sampler for detection of Bacillus endospores in indoor air
Authors:Rafał Lewandowski  Krystyna Kozłowska  Małgorzata Szpakowska  Elżbieta A Trafny
Institution:1. Department of Microbiology, Military Institute of Hygiene and Epidemiology, Kozielska 4, 01-163, Warsaw, Poland
Abstract:This study was designed to evaluate the measuring range and lowest limit of detection of Bacillus endospores in the ambient room air when the Sartorius MD8 sampler, and two different culture methods for bacterial enumeration were used. Different concentrations of bioaerosol were generated inside the test chamber filled with either the high-efficiency particulate air (HEPA)-filtered air or with the ambient room air. The detection of endospores in the HEPA-filtered air was achievable: (1) when they were aerosolized at a concentration above 7.56?×?103 CFU/m3 and analyzed with spread plate method, and (2) when they were aerosolized at a concentration above 4.00?×?102 CFU/m3 and analyzed with pour plate method. The detection of endospores in the ambient room air was possible: (1) when they were aerosolized at a concentration above 9.1?×?103 CFU/m3 and analyzed with spread plate method, and (2) when they were aerosolized at a concentration above 5.6?×?102 CFU/m3 and analyzed with pour plate method. The microorganisms present in the ambient room air interfere with precise quantification of Bacillus endospores when their concentration is relatively low. The results of this study may be helpful in critical assessment of the results obtained from monitoring the air for bacterial endospores.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号