影响地表水高锰酸盐指数自动监测数据准确性的因素
引用本文:姚志鹏,陈鹏,陈亚男,姜明岑,李晓明,刘允,王明翠,刘喜惠,刘京,杨凯.影响地表水高锰酸盐指数自动监测数据准确性的因素[J].中国环境监测,2022,38(5):203-208.
作者姓名:姚志鹏  陈鹏  陈亚男  姜明岑  李晓明  刘允  王明翠  刘喜惠  刘京  杨凯
作者单位:中国环境监测总站, 国家环境保护环境监测质量控制重点实验室, 北京 100012;江苏省环境监测中心, 江苏 南京 210019
基金项目:国家水体污染控制与治理科技重大专项(2017ZX07302002)
摘    要:地表水高锰酸盐指数自动监测仪器在实际应用中存在质控样测试和实际水样比对测试合格率相对偏低的问题。通过对高锰酸盐指数氧化还原机理的分析,结合目前国控水站对质控样品的测试数据,探讨了如何使用高锰酸盐指数自动监测仪分析测试得到准确可靠的水质监测数据。认为应尽量保持自动分析仪器的反应条件、稀释条件与标准方法一致,即取样量与试剂用量比与标准方法一致,并根据样品浓度变化实现量程自动切换,分量程进行仪器校准和测量,才能够提高自动分析仪器测量的准确性。

关 键 词:地表水|高锰酸盐指数|自动监测|仪器应用
收稿时间:2021/8/3 0:00:00
修稿时间:2022/5/30 0:00:00

Investigation of Factors Affecting the Accuracy of Permanganate Index Automatic Monitoring Data in Surface Water
YAO Zhipeng,CHEN Peng,CHEN Yanan,JIANG Mingcen,LI Xiaoming,LIU Yun,WANG Mingcui,LIU Xihui,LIU Jing,YANG Kai.Investigation of Factors Affecting the Accuracy of Permanganate Index Automatic Monitoring Data in Surface Water[J].Environmental Monitoring in China,2022,38(5):203-208.
Authors:YAO Zhipeng  CHEN Peng  CHEN Yanan  JIANG Mingcen  LI Xiaoming  LIU Yun  WANG Mingcui  LIU Xihui  LIU Jing  YANG Kai
Institution:State Environmental Protection Key Laboratory of Quality Control in Environmental Monitoring, China National Environmental Monitoring Centre, Beijing 100012, China;Jiangsu Province Environmental Monitoring Centre, Nanjing 210019, China
Abstract:The relatively low passing rate in comparison between quality control sample test and real water sample test troubled the automatic monitoring instrument of permanganate index for surface water in in practical application.Based on the analysis of the redox mechanism for permanganate index and the test data of quality control samples in the current state-controlled water monitoring station,this paper discusses how to get accurate and reliable water quality monitoring data using the automatic monitoring instrument of permanganate index.Keeping the reaction conditions and dilution conditions of automatic analysis instruments consistent with those of the standard method,i.e.,keeping the ratio of the sample quantity and reagent dosage consistent with that for the standard method,automatic switching of measurement ranges according to the sample concentrations,and calibrating the instrument in different measurement ranges,are key factors to improve the accuracy of automatic analysis instrument measurements.
Keywords:surface water|permanganate index|automatic monitoring|instrument application