大气预浓缩仪-GC/FPD测定环境空气中的痕量硫化物
引用本文:沈秀娥,常淼,刘保献,赵红帅,王小菊.大气预浓缩仪-GC/FPD测定环境空气中的痕量硫化物[J].中国环境监测,2015,31(6):103-108.
作者姓名:沈秀娥  常淼  刘保献  赵红帅  王小菊
作者单位:北京市环境保护监测中心, 北京 100048,北京市环境保护监测中心, 北京 100048,北京市环境保护监测中心, 北京 100048,北京市环境保护监测中心, 北京 100048,北京市环境保护监测中心, 北京 100048
摘    要:采用大气预浓缩仪-GC/FPD建立了环境空气中7种含硫化合物的测定方法。结果表明,经惰性化改造后的大气预浓缩仪-GC/FPD测定系统的基线低,对硫化物无残留。7种化合物的线性回归方程的相关系数均在0.995以上,高、中、低3种浓度的空白加标回收率分别为98.6%~105.3%、99.1%~103.7%、85.7%~99.7%,除硫化氢的相对标准偏差为10.9%外,其余6种含硫化合物的相对标准偏差可控制在6.9%以内。当进样体积为400 m L时,方法的检出限在0.10~1.1 ng/m3。CO2、H2O对实际样品的测定无干扰。因此,该方法能够满足目前监测工作的要求,更加适用于环境空气中痕量含硫化合物的测定。

关 键 词:含硫化合物  环境空气  大气预浓缩  气相色谱法  火焰光度检测器
收稿时间:2014/12/4 0:00:00
修稿时间:2015/1/30 0:00:00

Determination of Trace Sulfur Compounds in Environmental Air by GC/FPD Coupled the Air Preconcentrations
SHEN Xiu-e,CHANG Miao,LIU Bao-xian,ZHAO Hong-shuai and WANG Xiao-ju.Determination of Trace Sulfur Compounds in Environmental Air by GC/FPD Coupled the Air Preconcentrations[J].Environmental Monitoring in China,2015,31(6):103-108.
Authors:SHEN Xiu-e  CHANG Miao  LIU Bao-xian  ZHAO Hong-shuai and WANG Xiao-ju
Institution:Beijing Municipal Environmental Monitoring Centre, Beijing 100048, China,Beijing Municipal Environmental Monitoring Centre, Beijing 100048, China,Beijing Municipal Environmental Monitoring Centre, Beijing 100048, China,Beijing Municipal Environmental Monitoring Centre, Beijing 100048, China and Beijing Municipal Environmental Monitoring Centre, Beijing 100048, China
Abstract:GC/FPD coupled the Air Preconcentrations was applied to the determination of 7 kinds of trace sulfur compounds in environment air. It was found that the system blank of the Air Preconcentration-GC/FPD was clear, and no residuum after the analysis of sulfur compounds. Results showed that the correlation coefficient of 7 sulfur compounds were all above 0.995, it has high accuracy and precision. In addition, the recovery of sulfur compounds were in the range of 98.6%-105.3%,99.1%-103.7%,85.7%-99.7%. The relative standard deviations of the methods lower than 6.9% for 6 compounds, but for hydrogen sulfide, the relative standard deviation was 10.9%. Otherwise, when inject volume was 400 mL,the limit detection of 7 sulfur compounds was 0.10-1.1 ng/m3. So the measurement system was sensitive for sulfur compounds and very suitable for determination of the trace sulfur compounds in environment air.
Keywords:trace sulfur compounds  air preconcentrations  environment air  gas chromatography  flame photometric detector