基于氧猝灭的有机改性溶胶-凝胶微生物传感器测定BOD的研究
引用本文:戴媛静,钟振明,陈曦,王小如,KWOKYIN WONG.基于氧猝灭的有机改性溶胶-凝胶微生物传感器测定BOD的研究[J].环境科学学报,2003,23(5):683-688.
作者姓名:戴媛静  钟振明  陈曦  王小如  KWOKYIN WONG
作者单位:1. 厦门大学化学化工学院分析科学教育部重点实验室,厦门,361005
2. 香港理工大学应用生物与化学科技系,香港九龙
基金项目:国家 8 6 3计划环境与资源领域 2 0 0 1AA6 35 10 0课题,国家自然科学基金委员会科研基金资助项目,香港研究资助局科研基金资助项目
摘    要:通过有机改性溶胶凝胶PVA包埋方法固定微生物菌株,以氧光化学传感器作为二次传感,制备了响应良好的光化学BOD微生物传感器.在此基础上,考察了微生物传感膜的活化浓度、温度、pH、无机盐及部分金属离子对传感器响应的影响.并对传感器的重现性、储存稳定性和线性范围进行了考察,实验结果表明BOD含量在0—100mg/L浓度范围内与传感器的荧光值成线性关系,传感膜的响应活性保持较好,具有较长的使用寿命和良好的稳定性.

关 键 词:氧猝灭  有机改性溶胶-凝胶微生物传感器  测定  BOD  荧光值  响应活性  水体污染  水质监测
文章编号:0253-2468(2003)05-0683-06
收稿时间:2002/10/28 0:00:00
修稿时间:2002年10月28

Ormosils-PVA BOD sensing film based on oxygen luminescence quenching
DAI Yuanjing,ZHONG Zhenming,CHEN Xi,WANG Xiaoru and KWOKYIN WONG.Ormosils-PVA BOD sensing film based on oxygen luminescence quenching[J].Acta Scientiae Circumstantiae,2003,23(5):683-688.
Authors:DAI Yuanjing  ZHONG Zhenming  CHEN Xi  WANG Xiaoru and KWOKYIN WONG
Institution:The Key Laborary of Analytical Sciences of MOE and Department of Chemistry, Xiamen University, Xiamen 361005, China,The Key Laborary of Analytical Sciences of MOE and Department of Chemistry, Xiamen University, Xiamen 361005, China,The Key Laborary of Analytical Sciences of MOE and Department of Chemistry, Xiamen University, Xiamen 361005, China,The Key Laborary of Analytical Sciences of MOE and Department of Chemistry, Xiamen University, Xiamen 361005, China and Department of Applied Biology and Chemical Technology, The Hong Kong Polytechnic University, Hunghom, Kowloon, Hong Kong, China
Abstract:An organically modified silicate (ormosil)-PVA as an immobilization material for Hansennula anomala was investigated to develop an optochemical BOD sensor for the determination of BOD. After preconditioning, the BOD sensing film can be steadily used for two months. The linear fluctuant coefficient(R2) in the ranges of 0-100 mg/L and 0-40 mg/L was 0.9950 and 0.9604, respectively. The effects of activated concentration, temperature, pH, sodium chloride concentration and coexisting metal ions to BOD determination were investigated.
Keywords:Optochemical sensor  Microbe membrane  Ormosil  BOD determination