van der Graaf ER Limburg J Koomans RL Tijs M 《Journal of environmental radioactivity》2011,102(3):270-282
The calibration of scintillation detectors for gamma radiation in a well characterized setup can be transferred to other geometries using Monte Carlo simulations to account for the differences between the calibration and the other geometry. In this study a calibration facility was used that is constructed from bricks of well-known activity concentrations of 40K and of radionuclides from the 238U- and 232Th-series. Transfer of the calibration was attempted to a Marinelli beaker geometry with the detector inside a lead shield and to an in situ application with the detector positioned on a sand bed. In general this resulted in good correspondence (within 5-10%) between the activity concentrations derived using the transferred calibration and activities that were derived by independent measurements. Some discrepancies were identified that were attributed to coincident summing in the natural decay series and interference of radon. 相似文献