纳氏试剂两种配制方法对氨氮测定的影响比较
引用本文:徐立生,蔡裕丰.纳氏试剂两种配制方法对氨氮测定的影响比较[J].云南环境科学,2009,28(5):88-89.
作者姓名:徐立生  蔡裕丰
作者单位:启东市环境监测站,江苏,启东,226200 
摘    要:对纳氏试剂的两种配制方法在配制难易和含汞量、空白吸光度、校准曲线、检出限、精密度和加标回收率等方面作了比较,得出KI、HgCl2和KOH配制的纳氏试剂与KI、HgI2和NaOH配制的纳氏试剂相比,配制虽麻烦,但含汞量、空白吸光度和检出限均较低,而灵敏度、精密度和加标回收率却较高,可作为配制纳氏试剂的首选方法。而KI、HgI2和NaOH配制的纳氏试剂可用于应急监测中。

关 键 词:纳氏试剂  氨氮测定  配制方法  比较

Comparison of Ammonia Nitrogen Determination with Nessler Reagent by Preparing with Two Kinds of Different Methods
Institution:XU Li - sheng, CAI Yu - feng (Qidong Environmental Monitoring Station, Qidong Jiangsu 226200 China)
Abstract:Two different preparing methods of Nessler reagent used for ammonia nitrogen determination are compared on the aspects of ABS blank and calibration curve and determination limit and accuracy and recoveries of spiked samples. The results show that the Nessler reagent prepared by KI and HgCl2 and KOH is of low determination limit and ABS blank, but with high sensitivity and accuracy and recoveries of spiked samples although it is complicate to be prepared. The Nessler reagent prepared by KI and HgI2 and NaOH can be used in emergency monitoring.
Keywords:Nessler reagent  ammonia nitrogen determination  preparation method  compare