Determination of 226Ra at ultratrace level in mineral water samples by sector field inductively coupled plasma mass spectrometry
Authors:Zoriy Miroslav V  Varga Zoltan  Pickhardt Carola  Ostapczuk Peter  Hille Ralf  Halicz Ludwik  Segal Irina  Becker J Sabine
Institution:Central Division of Analytical Chemistry, Research Centre Juelich, D-52425, Juelich, Germany.
Abstract:An analytical procedure has been proposed for the determination of (226)Ra at the low femtogram per ml concentration level in mineral water samples using double focusing sector field ICP-MS (ICP-SFMS). For the pre-concentration and separation of radium from the matrix elements in water a tandem of a laboratory-prepared filter, based on MnO(2), and Eichrom "Sr-specific" resin was used. The recovery of the method was determined to be 70.5%. The limit of detection for (226)Ra determination was 0.02 fg ml(-1), including a pre-concentration factor of 10. In addition, uranium concentration and uranium isotope ratios were measured by ICP-SFMS. In several mineral water samples with a relatively high uranium content, (226)Ra concentrations were found between 0.7-15 fg ml(-1). The effective dose of the contribution was calculated using the radionuclide concentration and dose conversion factors from the World Health Organization, WHO (1993). Assuming a mineral water consumption of 2 l d(-1), a slightly higher calculated dose than the suggested limit for drinking water (0.1 mSv y(-1)) was found in some samples.
Keywords:
本文献已被 PubMed 等数据库收录!