首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Polar organic chemical integrative sampler (POCIS) uptake rates for 17 polar pesticides and degradation products: laboratory calibration
Authors:Imtiaz Ibrahim  Anne Togola  Catherine Gonzalez
Institution:1. Laboratory Division, Bureau de recherche géologiques et minières (BRGM), 3 avenue Claude Guillemin, 45100, Orléans, France
2. LGEI Center, Ecole des mines d’Alès (EMA), 6 Avenue de Clavieres, 30319, Alès, France
Abstract:Polar organic chemical integrative samplers (POCIS) are useful for monitoring a wide range of chemicals, including polar pesticides, in water bodies. However, few calibration data are available, which limits the use of these samplers for time-weighted average concentration measurements in an aquatic medium. This work deals with the laboratory calibration of the pharmaceutical configuration of a polar organic chemical integrative sampler (pharm-POCIS) for calculating the sampling rates of 17 polar pesticides (1.15?≤?logK ow?≤?3.71) commonly found in water. The experiment, conducted for 21 days in a continuous water flow-through exposure system, showed an integrative accumulation of all studied pesticides for 15 days. Three compounds (metalaxyl, azoxystrobine, and terbuthylazine) remained integrative for the 21-day experiment. The sampling rates measured ranged from 67.9 to 279 mL?day?1 and increased with the hydrophobicity of the pesticides until reaching a plateau where no significant variation in sampling rate is observed when increasing the hydrophobicity.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号