Assessment of opacimeter calibration according to International Standard Organization 10155
Authors:Gomes J F
Affiliation:Centro de Tecnologias Ambientais, Instituto de Soldadura e Qualidade, Oeiras Codex, Portugal.
Abstract:This paper compares the calibration method for opacimeters issued by the International Standard Organization (ISO) 10155 with the manual reference method for determination of dust content in stack gases. ISO 10155 requires at least nine operational measurements, corresponding to three operational measurements per each dust emission range within the stack. The procedure is assessed by comparison with previous calibration methods for opacimeters using only two operational measurements from a set of measurements made at stacks from pulp mills. The results show that even if the international standard for opacimeter calibration requires that the calibration curve is to be obtained using 3 x 3 points, a calibration curve derived using 3 points could be, at times, acceptable in statistical terms, provided that the amplitude of individual measurements is low.