Temperature calibration formula for activated charcoal radon collectors
Authors:Cooper Alexandre  Le Thiem Ngoc  Iimoto Takeshi  Kosako Toshiso
Institution:a Graduate School of Engineering, The University of Tokyo, 2-11-16 Yayoi, Bunkyo-ku, Tokyo 113-0032, Japan
b Institute of Nuclear Science and Technology, Vietnam Atomic Energy Commission, 59 Ly Thuong Kiet, Hanoi, Viet Nam
Abstract:Radon adsorption by activated charcoal collectors such as PicoRad radon detectors is known to be largely affected by temperature and relative humidity. Quantitative models are, however, still needed for accurate radon estimation in a variable environment. Here we introduce a temperature calibration formula based on the gas adsorption theory to evaluate the radon concentration in air from the average temperature, collection time, and liquid scintillation count rate. On the basis of calibration experiments done by using the 25 m3 radon chamber available at the National Institute of Radiological Sciences in Japan, we found that the radon adsorption efficiency may vary up to a factor of two for temperatures typical of indoor conditions. We expect our results to be useful for establishing standardized protocols for optimized radon assessment in dwellings and workplaces.
Keywords:Radon gas  PicoRad detectors  Activated charcoal collectors  Calibration experiment  Conversion factor  Temperature effect
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号