首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of 14 amines in air samples using midget impingers sampling followed by analysis with ion chromatography in tandem with mass spectrometry
Authors:Verriele Marie  Plaisance Hervé  Depelchin Laurence  Benchabane Samia  Locoge Nadine  Meunier Guillaume
Institution:Université Lille Nord de France, F-59000 Lille, France.
Abstract:An Ion Chromatography-Mass Spectrometry (IC-MS) method was developed for the simultaneous quantification of 14 volatile amines in air. The method includes collection of compounds into two midget impingers in a row filled with 15 ml of ultrapure water. The analytical performances with mass spectrometry detection were compared to those obtained with classical conductivity detection. The use of mass spectrometry detection (in SIM mode) overcomes most of the coelutions encountered with conductivity detection. Although the linearity domain of calibrations is reduced for the MS detection as compared with the CD detection, the detection limits in MS detection are highly lowered allowing the quantification of amines at the levels of μg m(-3) in air with a good accuracy for most compounds (RSD of less than 10%). This method was successfully applied to the analysis of amines released from polyurethane foams. Seven amines were identified and some in high concentrations, like dimethylaminoethanol, NIAX and TEDA.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏