首页 | 本学科首页   官方微博 | 高级检索  
     检索      

水质铜测定不确定度的评定
引用本文:赵红叶.水质铜测定不确定度的评定[J].环境科学与管理,2007,32(7):147-148,151.
作者姓名:赵红叶
作者单位:高淳县环境监测站,江苏,南京,211300
摘    要:根据火焰原子吸收分光光度法测定水中的铜含量,分析主要的测量不确定度来源,即标准曲线不确定度、标准溶液不确定度、测量重复性不确定度.计算得到水中铜的测定结果的合成不确定度为0.098mg/L,扩展不确定度为0.196mg/L.

关 键 词:  不确定度  评定
文章编号:1673-1212(2007)07-0147-02
修稿时间:2006-12-02

Evaluation of the Uncertainty for the Measurement of Copper in Water
Zhao Hongye.Evaluation of the Uncertainty for the Measurement of Copper in Water[J].Environmental Science and Management,2007,32(7):147-148,151.
Authors:Zhao Hongye
Institution:Gaochun Environmental Monitoring Station, Nanjing 211300, China
Abstract:The uncertainty of measurement for the determination of copper in water by Flame Atomic Absorption Spectrometry was evaluated.Many most possible sources of uncertainty of measurement were analyzed,they are :the uncertainty of standard curve,the uncertainty of standard aqua and the uncertainty of standard aqua.The result of repeated tests were that the combined uncertainty was 0.098mg/L and the expanded uncertainty was 0.196mg/L.
Keywords:copper  uncertainty  evaluation
本文献已被 CNKI 维普 万方数据 等数据库收录!