离子色谱法测定工作场所空气中的草铵膦 |
| |
引用本文: | 凌映茹,霍宗利,张锋,张昊,曹伊楠,王鑫楠,朱宝立. 离子色谱法测定工作场所空气中的草铵膦[J]. 环境监测管理与技术, 2024, 0(1): 57-59 |
| |
作者姓名: | 凌映茹 霍宗利 张锋 张昊 曹伊楠 王鑫楠 朱宝立 |
| |
作者单位: | 南京医科大学公共卫生学院;江苏省疾病预防控制中心 |
| |
基金项目: | 江苏省医学重点学科基金资助项目(ZDXK202249) |
| |
摘 要: | 用玻璃纤维滤纸采集工作场所空气中的草铵膦,经Metrosep A Supp 5-150型阴离子色谱柱分离,以1.92mmol/L碳酸钠-0.60mmol/L碳酸氢钠混合溶液为流动相,用离子色谱法测定。该方法在0mg/L~1.04mg/L范围内线性良好,方法检出限为0.012mg/L,按采集30 L空气计算,最低检出值为0.004mg/m3。空白玻璃纤维滤纸的加标回收率为93.7%~102%,标准溶液6次测定结果的RSD为0.8%~4.3%。
|
关 键 词: | 草铵膦;离子色谱法;工作场所;环境空气 |
Determination of Glufosinate in the Air of Workplace by Ion Chromatography |
| |
Abstract: | Glufosinate in the air of workplace was collected by glass fiber filter paper, separated by Metrosep A Supp 5-150 anion chromatographic column and determined by ion chromatography, with 1.92 mmol/L sodium carbonate-0.60 mmol/L sodium bicarbonate mixed solution as mobile phase. The method had a good linearity in the range of 0 mg/L~1.04 mg/L. The method detection limit was 0.012 mg/L. The minimum detectable limit was 0.004 mg/m3 when the collection volume was 30 L. The recoveries of spiked blank samples in glass fiber filter paper were from 93.7% to 102%,and the RSDs of six measurements of the standard solutions were from 0.8% to 4.3%. |
| |
Keywords: | Glufosinate Ion chromatography Workplace Ambient air |
|