Automated indirect fluoride determination by ICP-OES using discontinuous-flow analysis
Authors:Máté Kovács  Riccardo Magarini  Pál Halmos
Institution:1. Institute of Analytics, Environmental Science and Limnology, University of Pannonia , Veszprém , Hungary mkovacs@almos.uni-pannon.hu;3. PerkinElmer Life and Analytical Sciences , Monza , 20052 , Italy;4. Institute of Analytics, Environmental Science and Limnology, Hungarian Academy of Sciences, University of Pannonia , Veszprém , Hungary
Abstract:An automated method for the indirect determination of fluoride by inductively coupled plasma atomic emission spectrometry using discontinuous-flow analysis has been developed. A modified flow injection analysis system was used for sample preparation. The method is based on the determination of excess cerium(III) ion after precipitation of CeF3. Separation of the precipitate is performed online on a syringe membrane filter that is cleaned each analysis cycle by 1 M nitric acid, ensuring a long life-time for the filter. As a result of the short reaction time, a detection limit of 9.5 mg L?1 was obtained. The precision (RSD) of the proposed method is 1.02% at fluoride concentration of 28 mg L?1. The applicability of the technique for groundwater monitoring of fluoride-contaminated area is demonstrated.
Keywords:fluoride  inductively coupled plasma optical emission spectrometry (ICP-OES)  cerium(III) fluoride  precipitation  automation