首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An inexpensive light-scattering particle monitor: field validation
Authors:Chowdhury Zohir  Edwards Rufus D  Johnson Michael  Naumoff Shields Kyra  Allen Tracy  Canuz Eduardo  Smith Kirk R
Institution:School of Public Health, University of California at Berkeley, California, USA. zohir@berkeley.edu
Abstract:We have developed a small, light, passive, inexpensive, datalogging particle monitor called the "UCB" (University of California Berkeley particle monitor). Following previously published laboratory assessments, we present here results of tests of its performance in field settings at high particle concentrations. We demonstrate the mass sensitivity of the UCB in relation to gravimetric filter-based PM(2.5) mass estimates as well as commercial light-scattering instruments co-located in field chamber tests and in kitchens of wood-burning households. The coefficient of variation of the unadjusted UCB mass response in relation to gravimetric estimates was 15%. Although requiring adjustment for differences in sensitivity, inter-monitor performance was consistently high (r(2) > 0.99). Moreover, the UCB can consistently estimate PM(2.5) mass concentrations in wood-burning kitchens (Pearson r(2) = 0.89; N = 99), with good agreement between duplicate measures (Pearson r(2) = 0.94; N = 88). In addition, with appropriate cleaning of the sensing chamber, UCB mass sensitivity does not decrease with time when used intensively in open woodfire kitchens, demonstrating the significant potential of this monitor.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号