首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of a fluorometric method for measuring low concentrations of ammonia in ambient air
Authors:Groves William A  Agarwal Divya  Chandra M Jeya  Reynolds Stephen J
Institution:Industrial Health and Safety Program, Department of Energy and Geo-Environmental Engineering, Penn State University, University Park, PA 16802, USA.
Abstract:A fluorometric method developed for measuring low concentrations of ammonium in marine and freshwater ecosystems was adapted for the analysis of ammonia in ambient air. The modified method entails collection of samples on an acid-treated solid adsorbent followed by analysis using a fluorometer. Optimal results were obtained using a commercially available sorbent tube containing 100 mg of acid-treated silica gel for sample collection, and an analytical protocol consisting of sample desorption in DI water, addition of orthopthaldialdehyde (OPA) working reagent, and room temperature incubation. Method accuracy and precision were evaluated by comparing experimentally determined quantities of ammonia to expected levels for sample loadings ranging from 0.16 micro sign]g to 550 micro sign]g-accuracy was generally within +/-20%. The estimated LOQ for the method is 0.08 micro sign]g ammonia per sample which represents a 25-375-fold improvement in sensitivity compared to current NIOSH and OSHA methods for the measurement of ammonia in ambient air. The new method should be useful for applications requiring measurement of low concentrations of ammonia using personal sampling equipment or in the characterization of short-term fluctuations of ammonia concentrations in air.
Keywords:
本文献已被 PubMed 等数据库收录!