Development of a diffusive sampling method for determination of methyl isocyanate in air
Authors:von Zweigbergk Peter  Lindahl Roger  Ostin Anders  Ekman Jenny  Levin Jan-Olof
Institution:National Institute for Working Life, Ume?, Sweden. peter.von.zweigbergk@niwl.se
Abstract:A diffusive sampling method for determination of methyl isocyanate in air has been developed. A glass fibre filter impregnated with 1-(2-methoxyphenyl)piperazine in a commercially available diffusive sampling device was used to collect methyl isocyanate and the derivative formed was analysed with LC-MS/MS. The sampling rate was determined to be 15.6 ml min(-1), with a relative standard deviation of 7.3%. The sampler was validated for sampling periods from 15 min to 8 h, for relative humidities from 20% to 80% and for concentrations from I to 46 microg m(-3). A field validation was also made and the diffusive sampling results showed no difference compared to a pumped reference method. The impregnated filters have to be stored apart from the diffusive sampler housing and loaded into the sampler prior to each sampling.