首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of response of real-time SidePak AM510 monitor to secondhand smoke, other common indoor aerosols, and outdoor aerosol
Authors:Jiang Ruo-Ting  Acevedo-Bolton Viviana  Cheng Kai-Chung  Klepeis Neil E  Ott Wayne R  Hildemann Lynn M
Affiliation:Civil & Environmental Engineering Dept, Stanford University, Stanford, CA 94305, USA.
Abstract:The amount of light scattered by airborne particles inside an aerosol photometer will vary not only with the mass concentration, but also with particle properties such as size, shape, and composition. This study conducted controlled experiments to compare the measurements of a real-time photometer, the SidePak AM510 monitor (SidePak), with gravimetric mass. PM sources tested were outdoor aerosols, and four indoor combustion sources: cigarettes, incense, wood chips, and toasting bread. The calibration factor for rescaling the SidePak measurements to agree with gravimetric mass was similar for the cigarette and incense sources, but different for burning wood chips and toasting bread. The calibration factors for ambient urban aerosols differed substantially from day to day, due to variations in the sources and composition of outdoor PM. A field evaluation inside a casino with active smokers yielded calibration factors consistent with those obtained in the controlled experiments with cigarette smoke.
Keywords:
本文献已被 PubMed 等数据库收录!