首页 | 本学科首页   官方微博 | 高级检索  
     


Addition of surrogate standards in the routine analysis of investigating polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/F)
Authors:Lars-Owe Kjeller
Affiliation:1. Institute of Environmental Chemistry, Ume? University, S-901 87, Ume?, Sweden
Abstract:Conclusion  Independent losses of internal standard by evaporation and/or sorption will increase both absolute and random errors for the native analyte. Thus, the internal standard should only be used for compensation of losses, variations and irregularities in the clean-up, dilution, transfer and instrument. This is achieved by addition of internal standard to the eluent after extraction, or avoiding direct contact between internal standards and suspected active matrices. The analytical results will then not guarantee a correct quantification of the true analyte concentration on the matrix, only maximise the precision of measurement of extracted analyte. In Soxhlet extraction, the internal standard can be added to the solvent reservoir. In supercritical fluid extraction (SFE), the internal standard might be added to an aliquot of reversed phase silica (C-8, C-18) located after the sample in the fluid direction.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号