首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An improved dual channel PERCA instrument for atmospheric measurements of peroxy radicals
Authors:Green Timothy J  Reeves Claire E  Fleming Zoe L  Brough Neil  Rickard Andrew R  Bandy Brian J  Monks Paul S  Penkett Stuart A
Institution:School of Environmental Science, University of East Anglia, University Plain, Norwich, UKNR4 7TJ.
Abstract:This paper describes a new dual-channel PEroxy RadiCal Amplification (PERCA) instrument, which has been designed to improve the time resolution and signal to noise and to reduce the interference caused by variations in ambient ozone concentrations. The instrument was run at the Weybourne Atmospheric Observatory (WAO), North Norfolk, during WAOWEX (Weybourne Atmospheric Observatory Winter Experiment) in January/February 2002 and INSPECTRO (Influence of clouds on the spectral actinic flux in the lower troposphere) in September 2002. The performance of the instrument is assessed and compared to that of a single channel instrument. In particular, it is shown how the precision is greatly improved in fluctuating background ozone conditions. In addition the improved time response of the instrument allows changes in peroxy radical concentrations to be related to rapid changes in nitric oxide concentrations and the ozone photolysis frequency, j(O(1)D).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏