电感耦台等离子体发射光谱法测定锅炉水中硫、磷
引用本文:王义壮,王菁婷,韦中悬.电感耦台等离子体发射光谱法测定锅炉水中硫、磷[J].中国特种设备安全,2014(8):50-52.
作者姓名:王义壮  王菁婷  韦中悬
作者单位:广东省特种设备检测研究院东莞检测院,东莞525416
摘    要:本文对利用电感耦合等离子体发射光谱法(ICP--AES)测定锅炉水中硫、磷含量的方法进行了探讨。对射频功率、泵速、雾化器流量和观察高度等仪器条件进行了优化选择,考察了检测方法的准确度和精密度。硫、磷的方法检出限分别为0.035mg/L、0.089mg/L,相对标准偏差(RSD,n=9)为0.52%~0.69%。

关 键 词:ICP--AES  锅炉    

Sulphur and Phosphorous Measured by Inductively Coupled Plasma Atomic Emission Spectrometry in boiler water
Authors:Wang Yizhuang Wang Jingting Wei Zhongxuan
Institution:Wang Yizhuang Wang Jingting Wei Zhongxuan (Guangdong Province Special Equipment Inspection Institute Branch of Dongguan Dongguan 523416)
Abstract:The inductively coupled plasma atomic emission spectrometry (ICP-AES) method was developed for determination of S, P in boiler water. The instrument operating parameters were optimized such as power, nebulizer flow, pump rate, observation height. The detection accuracy and precision of the method were investigated. The detection limit of the method was 0.035 mg / L for SO4 2-, 0.089rag/L for PO4 3-. The precision of the method RSD n=9 were between 0.32%-0.69%.
Keywords:ICP-AES  Boiler  Sulphur  Phosphorus