首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Validation of atmospheric aerosols parallel sampling in a multifold device
Authors:C M Oliveira  M F Camões  P Bigus  A A Fachado  R B Silva
Institution:1.Centro de Química Estrutural—Faculdade de Ciências,Universidade de Lisboa,Lisboa,Portugal;2.Department of Analytical Chemistry, Chemical Faculty,Gdańsk University of Technology,Gdańsk,Poland
Abstract:In this work, particulate matter was collected using an active sampling system consisting of a PM10 (<10 μm) inlet coupled to a multifold device containing six channels, connected to a vacuum pump. Each channel was equipped with a filter holder fitted with adequately chosen filters. The system was fixed on a metallic structure, which was placed on the roof of the laboratory building, at the Faculty of Sciences, in Lisbon. Sampling took place under flow-controlled conditions. Aerosols were extracted from the filters with water, in defined conditions, and the water-soluble fraction was quantified by ion chromatography (IC) for the determination of inorganic anions (Cl?, NO3 ? and SO4 2?). Equivalent sampling through the various channels was validated. Validation was based on the metrological compatibility of the content results for the various filters. Ion masses are metrologically equivalent when their absolute difference is smaller than the respective expanded uncertainty. When this condition is verified, the studied multifold device produces equivalent samples.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏