首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Speciation of arsenic using solid phase extraction cartridges
Authors:Yalçin S  Le X C
Institution:Environmental Health Sciences Program, Department of Public Health Sciences, Faculty of Medicine, University of Alberta, Edmonton, Canada.
Abstract:Various solid phase extraction (SPE) cartridges were investigated for speciation of arsenite As(III)], arsenate As(v)], monomethylarsonic acid (MMA) and dimethylarsinic acid (DMA). Cartridges containing different types of sorbent materials were tested for arsenic retention and elution characteristics. Alumina cartridges were found to completely retain all the four target arsenic species, and are suitable for removal and preconcentration purposes. For speciation analysis, different arsenic species were separated on the basis of their selective retention on and elution from specific cartridges. DMA was retained on a resin-based strong cation exchange cartridge and eluted with 1.0 M HCl. MMA and As(v) were both retained on a silica-based strong anion exchange cartridge and sequentially eluted with 60 mM acetic acid (for MMA) and 1.0 M HCl for As(v)]. As(III) was not retained on either cartridge and remained in solution. Arsenic species in solution and those eluted from the cartridges were subsequently quantified by using flow injection with hydride generation atomic fluorescence spectrometry (FI-HGAFS) and hydride generation atomic absorption spectrometry (FI-HGAAS). A detection limit of 0.05 microg L(-1) arsenic in water sample was achieved using HGAFS. An application of the method was demonstrated at a drinking water treatment facility. As(III) and As(v) species were determined in water at various stages of treatment. The method is suitable for routine determination of trace levels of arsenic in drinking water to comply with more stringent environmental regulations.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号