Determination of hydrogen peroxide in workplace air: interferences and method validation
Authors:Christensen C S  Brødsgaard S  Mortensen P  Egmose K  Linde A
Institution:MILJ?-KEMI, Danish Environmental Center A / S, Galten.
Abstract:A dynamic system for the generation of stable hydrogen peroxide test atmospheres was applied to the evaluation of samplers used for the determination of hydrogen peroxide in workplace air. The system is able to generate gas mixtures of between 0.1 and 10 ppm at different combinations of relative humidity (20-80%) and temperature (10-30 degrees C). Gaseous hydrogen peroxide is sampled on glass filters impregnated with Ti(IV) chloride and sulfuric acid and analyzed by UV spectroscopy. An interference was observed due to the acid catalyzed decomposition of peroxyacetic acid to hydrogen peroxide. This was significantly reduced by applying high sample flow rates and by lowering the concentration of sulfuric acid. The performance of the sampler and sources of uncertainties were tested according to the European Standard EN 1076.