首页 | 本学科首页   官方微博 | 高级检索  
     检索      

气袋直接进样法测定丙烯腈
引用本文:李冰清.气袋直接进样法测定丙烯腈[J].环境保护科学,2012,38(4):70-72.
作者姓名:李冰清
作者单位:上海市环境监测中心,上海,200030
摘    要:采用气袋直接进样法测定环境和污染源空气中的丙烯腈。结果表明:丙烯腈在0.4~99.5 mg/m3浓度范围内具有良好的线性关系,相关系数为0.999 8。在低、中、高3个添加水平回收率为88.9%~100%,相对标准偏差为0.6%~4.7%,均能较好地满足实验室分析要求。

关 键 词:气袋  直接进样  丙烯腈  气相色谱

Determination of Acrylonitrile by Direct Injection into Airbag
Abstract:Acrylonitrile in air environment and air pollution sources was determined by direct injection into airbag. The results showed that calibration curve of acrylonitrile in concentration range of 0.4-99.5 mg/m^3 was linear, and the correlation coefficient was 0.999 8. The recoveries at three levels of low, middle and high were in the range of 88.9%-100%, relative standard deviation were in the range of 0.6%-4.7%, which could met the requirements of lab analysis.
Keywords:Airbag  Direct Injection  Acrylonitrile  Gas Chromatography