首页 | 本学科首页   官方微博 | 高级检索  
     检索      

TAR柱前络合、毛细管电泳分析痕量金属离子
引用本文:王敏,屈锋,林金明.TAR柱前络合、毛细管电泳分析痕量金属离子[J].环境化学,2003,22(3):236-240.
作者姓名:王敏  屈锋  林金明
作者单位:中国科学院生态环境研究中心,北京,100085
基金项目:SupportedbyNSFC (No. 2 0 1 2 5 5 1 4andNo . 2 0 2 75 0 43 )
摘    要:1 IntroductionTransitionmetalswereamongthemostconcernelementsinenvironmentalanalysisduetotheirpotentialrolestoplantandanimals1 ].Theserolesoftransitionmetalsinenvironmentdependedontheirconcentrationinthesoil,waterandlivingthings.Currently ,themostutiliz…

关 键 词:TAR柱  柱前络合  毛细管电泳分析  痕量分析  金属离子

CAPILLARY ELECTROPHORESIS FOR THE ANALYSIS OF TRACE METAL IONS USING PRE-COLUMN COMPLEXING WITH 4-(2-THIAZOLYLAZO)RESORCINOL
Abstract.CAPILLARY ELECTROPHORESIS FOR THE ANALYSIS OF TRACE METAL IONS USING PRE-COLUMN COMPLEXING WITH 4-(2-THIAZOLYLAZO)RESORCINOL[J].Environmental Chemistry,2003,22(3):236-240.
Authors:Abstract
Abstract:The capillary electrophoretic separation of Fe2+, Co2+, Zn2+ and Ni2+ in a phosphate buffer solution by complexing with 4-(2-thiazolylazo)resorcinol was investigated. The influences of some crucial parameters that included chelating ligand in the electrophoretic running buffer and sample solution, pH value and concentration of buffer were examined. Under optimum conditions (10mmol*l-1 NaH2PO4-Na2HPO4 buffer containing 1×10-4mol*l-1TAR, pH 7.5), a baseline separation of these metals was accomplished within 3 min. The detection limits (S/N=3) ranged from 0.013-0.14 μg*ml-1. The method was applied to analyze trace metal ions in the environmental samples.
Keywords:capillary zone electrophoresis  metal ions  4  (2  thiazolylazo)resorcinol    
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏